LEAHI-TD-FIRMWARE-LDT-4027_Pre-Treatment - FW Implementation - 2/6:
LEAHI-TD-FIRMWARE-LDT-4027_Pre-Treatment - FW Implementation - 2/6:
It looks like you have this as a T/F response from UI. UI should not be determining whether the modality matches - that's our job. UI should be telling us which type of tubing set was installed (HD...

It looks like you have this as a T/F response from UI. UI should not be determining whether the modality matches - that's our job. UI should be telling us which type of tubing set was installed (HD or HDF type). Then when we get this response msg from UI, we will compare the type to the user selected modality (from TxParams) to see if it matches.

Add doxygen comments to the right of each structure field.

Add doxygen comments to the right of each structure field.

Why is this commented out? We will get stuck in this state if we don't move on to the pressure leak test states.

Why is this commented out? We will get stuck in this state if we don't move on to the pressure leak test states.

For now, we need to set authResponseReceived and authResponseValidTubingSet and authResponseModalityAccepted to TRUE until UI is properly responding to authentication request (TODO).

For now, we need to set authResponseReceived and authResponseValidTubingSet and authResponseModalityAccepted to TRUE until UI is properly responding to authentication request (TODO).

For now, we need to set barcodeScanResponseReceived to TRUE here until barcode scan command is implemented.

For now, we need to set barcodeScanResponseReceived to TRUE here until barcode scan command is implemented.

Add blank line before return statement.

Add blank line before return statement.

Keep this blank line above group header.

Keep this blank line above group header.

Something is wrong here. We already had reg 33 at line 192 above. Reconcile with HDD.

Something is wrong here. We already had reg 33 at line 192 above. Reconcile with HDD.

Make sure UI s/w is looking for this message on the broadcast channel or change this channel to the TD=>UI channel if UI is looking for this message on that channel.

Make sure UI s/w is looking for this message on the broadcast channel or change this channel to the TD=>UI channel if UI is looking for this message on that channel.

where are we calling an alarm? i think we are only sending event change data in completeBolusToCumulative() right?

where are we calling an alarm?
i think we are only sending event change data in completeBolusToCumulative() right?

do we need this block?

do we need this block?

update the comment as required: ml or L?

update the comment as required: ml or L?

rejReason = INVALID REQUEST FORMAT

rejReason = INVALID REQUEST FORMAT

I think these remaining 4 else ifs need to have another && condition that your service is in idle state so that we don't change back into one of these 4 states while a bolus is in progress.

I think these remaining 4 else ifs need to have another && condition that your service is in idle state so that we don't change back into one of these 4 states while a bolus is in progress.

Can we restore the real code here yet? Ask Dara.

Can we restore the real code here yet? Ask Dara.

Why isn't this TRUE == ?

Why isn't this TRUE == ?

LEAHI-TD-FIRMWARE-LDT-3975_Water Treatment and Degassing - FW Iteration - 4/4:
LEAHI-TD-FIRMWARE-LDT-3975_Water Treatment and Degassing - FW Iteration - 4/4:
use ternary to make TRUE or FALSE.

use ternary to make TRUE or FALSE.

Fixed

Fixed

Yes

Yes

REmoved

REmoved

Moved

Moved

Fixed

Fixed

Removed

Removed

Not a float.

Not a float.

Not a float.

Not a float.