MixingControl.c

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
There are multiple target bicarb conductivity calls, getBicarbTargetConductivity, getBicardConductivityPre and getBicarConductivityPost. If you are using getBicarbTargetConductivity for deadband co...

There are multiple target bicarb conductivity calls, getBicarbTargetConductivity, getBicardConductivityPre and getBicarConductivityPost. If you are using getBicarbTargetConductivity for deadband control check, use the same for 'runPIcontroller' function as well.

This should return POST conductivity.

This should return POST conductivity.

getBicarbConductivityPre() -> This function gets the target conductivity from the mixing algorithm. For testing, recommend using defined value and check the behavior.

getBicarbConductivityPre() -> This function gets the target conductivity from the mixing algorithm. For testing, recommend using defined value and check the behavior.

Init PI Controls missing here.

Init PI Controls missing here.

13734.87F

13734.87F

Merge branch LDT-4612-dialysate-composition-integ into develop

LDT-4612 modifying the acid type as 1K

LDT-4612 mixing control updates

LDT-4612 mixing control splitting updates

Merge branch 'LDT-4410-dialysate-composition-mixing-2' into develop

LDT-4410 addressed code review comments

  1. … 1 more file in changeset.
Can we move this exec at task level, independent of Drybicart operations?

Can we move this exec at task level, independent of Drybicart operations?

Comment missing?

Comment missing?

Leave a comment here that for acid control period, 3 cycles of balancing chamber switching time (for ex, Qd :600, 3sec is switching period, 3 * 3 = 9 Sec control interval) and for bicarb control, 5...

Leave a comment here that for acid control period, 3 cycles of balancing chamber switching time (for ex, Qd :600, 3sec is switching period, 3 * 3 = 9 Sec control interval) and for bicarb control, 5 cycles of balancing chamber switching time.

both acid and bicarb mix volume

both acid and bicarb mix volume

Add MULTIPLIER at the end.

Add MULTIPLIER at the end.

Initialize at init function.

Initialize at init function.

Remove sample filter ms and task interval. rename as FILTER_SIZE_MULTIPLIER.

Remove sample filter ms and task interval. rename as FILTER_SIZE_MULTIPLIER.

Lets remove this.

Lets remove this.

LEAHI-DD-FIRMWARE-LDT-4410_Dialysate Composition (Mixing) - FW Implementation - 3/1:
LEAHI-DD-FIRMWARE-LDT-4410_Dialysate Composition (Mixing) - FW Implementation - 3/1:
LDT-4410 inital code for filter and control interval

  1. … 2 more files in changeset.
LDT-4612 fix for mixing control

  1. … 1 more file in changeset.
LDT-4409 fix for concentrate pump not running

  1. … 2 more files in changeset.
LEAHI-DD-FIRMWARE-LDT-4612_Dialysate Composition (Dry Bicarbonate) - FW Iteration - 4/1:
LEAHI-DD-FIRMWARE-LDT-4612_Dialysate Composition (Dry Bicarbonate) - FW Iteration - 4/1:
LDT-4612 fix for micing control integration issues

  1. … 1 more file in changeset.
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 13 more files in changeset.
LDT-4409 addressed review comments

  1. … 3 more files in changeset.