Remove extra blank line.

Remove extra blank line.

It is your comment, so please resolve it.

It is your comment, so please resolve it.

Merged staging

It is replied by Vinay , there is no need to maintain two different rpm. so we can close this.

It is replied by Vinay , there is no need to maintain two different rpm. so we can close this.

It will be fixed in Balancing chamber re-implementation

It will be fixed in Balancing chamber re-implementation

I would prefer to fix this in balancing chamber new algorithm implementation. That is why i didn't mark it as Unresolved.

I would prefer to fix this in balancing chamber new algorithm implementation. That is why i didn't mark it as Unresolved.

input parameter not updated

input parameter not updated

input parameter is not updated

input parameter is not updated

input and output is not updated properly

input and output is not updated properly

structure packing required

structure packing required

fixed it

fixed it

fixed it

fixed it

This state machine running continuously. fill request is initiated for every 1 minutes of balancing chamber exec (since right now level sensor is not working ) //Increment counter for dry bicarb c...

This state machine running continuously. fill request is initiated for every 1 minutes of balancing chamber exec (since right now level sensor is not working )

//Increment counter for dry bicarb chamber fill
bicarbChamberPeriodicFillCounter += 1;
// Fill bicarb chamber once every 60secs.
if ( bicarbChamberPeriodicFillCounter >= BICARB_CHAMBER_PERIODIC_FILL_TIME )

Unknown macro: { setBicarbChamberFillRequested(); bicarbChamberPeriodicFillCounter = 0; }


The request is cleared here. for consistency I am moving to the end of the fill cycle

LEAHI-TD-FIRMWARE-LDT-1415_Pressures Monitoring - TD
LEAHI-TD-FIRMWARE-LDT-1415_Pressures Monitoring - TD
LDT-1415-pressures-monitoring-td

LDT-1415-pressures-monitoring-td

Merge branch 'develop' of bitbucket.org:diality-cloud/leahi-common into develop

Yes, we do and this is not required here so, I removed

Yes, we do and this is not required here so, I removed

LDT-1415-pressures-monitoring-td

Shouldn't we be getting these defaults from the TxParams unit?

Shouldn't we be getting these defaults from the TxParams unit?

LEAHI-DD-FIRMWARE-LDT-2004_Dialysate Composition - DD
LEAHI-DD-FIRMWARE-LDT-2004_Dialysate Composition - DD
Yes, Updated

Yes, Updated

Why is this flag set to FALSE here in start state while the other request flags are set to FALSE in their completed states?

Why is this flag set to FALSE here in start state while the other request flags are set to FALSE in their completed states?