dd-firmware

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
LDT-3980- adjusting bicarb kp gain

LDT-3749 merged into develop

    • -0
    • +80
    /firmware/App/Modes/ModeGenDialysate.c
LDT-3749 merged into staging

LDT-3749 resolved merge conflict with staging

    • -57
    • +58
    /firmware/App/Modes/ModeGenDialysate.c
    • -3
    • +3
    /firmware/App/Modes/ModeGenDialysate.h
    • -2
    • +5
    /firmware/App/Modes/ModePreGenDialysate.c
    • -14
    • +14
    /firmware/App/Modes/ModeStandby.c
    • -20
    • +22
    /firmware/App/Services/Messaging.c
Bamboo Commit: Updated DDCommon.h with build versions from Bamboo.

Merge branch 'LDT-3972-water-treatment-and-degassing-1' into develop

LDT-3972: changed filter size to be 2.1s from 3.5s for cond and cond temp.

Merge branch 'LDT-3972-water-treatment-and-degassing-1' into develop

LDT-3972: removal of dead code

Merge branch 'staging' into LDT-3972-water-treatment-and-degassing-1

Bamboo Commit: Updated DDCommon.h with build versions from Bamboo.

Merge branch 'LDT-3972-water-treatment-and-degassing-1' into develop

LDT-3972: fixed freshness boolean for conductivity sensors.

LDT-3223-dialysate-flow-rate---dd---mode-tel-dev-test-3-30-2026 : Added test cases for BloodLeak and Pressuresensor.c

LDT-3223-dialysate-flow-rate---dd---mode-tel-dev-test-3-30-2026 : Added test cases for Comm.c file

LDT-3735: receive in progress

    • -92
    • +169
    /firmware/App/Services/NVMsgQ.c
    • -13
    • +13
    /firmware/App/Services/NVRecordsDD.c
control interval is different. Control interval must be separated else we don't get finer control on bicarb conductivity. also the dead band is different. need to figure out whether we can combine ...

control interval is different. Control interval must be separated else we don't get finer control on bicarb conductivity. also the dead band is different. need to figure out whether we can combine or not

This code looks very similar to bicarb code above. Do we really need separate handling?

This code looks very similar to bicarb code above. Do we really need separate handling?

Updated

Updated

LDT-3980 firmware development for dry bicart updates

LDT-3980 firmware development for dry bicart

    • -32
    • +74
    /firmware/App/Controllers/DryBiCart.c
Remove extra blank line.

Remove extra blank line.

Add unit of measure to definition name (e.g. _PSI). Should be done for all definitions w/ a unit of measure to prevent unit errors.

Add unit of measure to definition name (e.g. _PSI). Should be done for all definitions w/ a unit of measure to prevent unit errors.

LEAHI-DD-FIRMWARE-LDT-3980_Dialysate Composition - FW Implementation (Dry Bicarbonate) - 1/1:
LEAHI-DD-FIRMWARE-LDT-3980_Dialysate Composition - FW Implementation (Dry Bicarbonate) - 1/1:
LDT-3980-Dry bicarb firmware implementation for initial fill request

    • -10
    • +25
    /firmware/App/Controllers/DryBiCart.c