we can start with 500msec. if we change the speed during runtime, say 10% to 20% PWM, need to see how long it takes to stabilize and accordingly this persistence can be set.

we can start with 500msec. if we change the speed during runtime, say 10% to 20% PWM, need to see how long it takes to stabilize and accordingly this persistence can be set.

fixed

fixed

reworded

reworded

Updated

Updated

LDT-3539-air-infusion-td

Based on the Denali properties, I have written the following.

Based on the Denali properties, I have written the following.

LDT-3539-air-infusion-td

replace with else block

replace with else block

convert to switch case

convert to switch case

set treatment modality.

set treatment modality.

Priority doesn't look right. Should be High I think. Check properties from Denali project.

Priority doesn't look right. Should be High I think. Check properties from Denali project.

LEAHI-TD-FIRMWARE-LDT-3539_Air Infusion - TD
LEAHI-TD-FIRMWARE-LDT-3539_Air Infusion - TD
Vinay, based on your work with rinsepump, Do you have any persistence value in mind?

Vinay, based on your work with rinsepump, Do you have any persistence value in mind?

LDT-3539-air-infusion-td

Can we check some persistence before declaring alarm?

Can we check some persistence before declaring alarm?

recommend running the pump at few sample speeds and observe the measured feedback and fix the tolerance number.

recommend running the pump at few sample speeds and observe the measured feedback and fix the tolerance number.

Rinse pump fault? do we have one? remove concentrate references, commented codes, empty lines.

Rinse pump fault? do we have one? remove concentrate references, commented codes, empty lines.

No, Yes for lowering event. Fill event uses PWM 150 always irrespective of Qb and Venous pressure only lower event used equation based PWM and that is in the range of 60-250 , not default., based ...

No, Yes for lowering event.

Fill event uses PWM 150 always irrespective of Qb and Venous pressure
only lower event used equation based PWM and that is in the range of 60-250 , not default., based on venous pressure it changes

Do we ever want to use the default (250)? Do we always want to use this venous pressure based power level?

Do we ever want to use the default (250)? Do we always want to use this venous pressure based power level?

Why not for release?

Why not for release?

Remove extra blank line.

Remove extra blank line.

Remove extra blank line.

Remove extra blank line.

Remove extra blank line.

Remove extra blank line.

LEAHI-TD-FIRMWARE-LDT-1903_Air Trap Level Control - TD - 04: DEV - Feature Implementation
LEAHI-TD-FIRMWARE-LDT-1903_Air Trap Level Control - TD - 04: DEV - Feature Implementation
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

Merged LDT-2749 air bubble alarm