instead of 0, we can use the level Enum 'LVL_STATE_EMPTY'?

instead of 0, we can use the level Enum 'LVL_STATE_EMPTY'?

lower? and change the D98 level sensor comment as well.

lower? and change the D98 level sensor comment as well.

Remove empty line

Remove empty line

Space before LVL_STATE_LOW

Space before LVL_STATE_LOW

space inside array index.

space inside array index.

Space inside parenthesis.

Space inside parenthesis.

Change this to LEVEL_T enum.

Change this to LEVEL_T enum.

Is this macro still being used? if not, delete it.

Is this macro still being used? if not, delete it.

LEAHI-DD-FIRMWARE-LDT-3978_Switch Chamber F conductive level sensors to broadcast 1 as sensing liquid
LEAHI-DD-FIRMWARE-LDT-3978_Switch Chamber F conductive level sensors to broadcast 1 as sensing liquid
control interval is different. Control interval must be separated else we don't get finer control on bicarb conductivity. also the dead band is different. need to figure out whether we can combine ...

control interval is different. Control interval must be separated else we don't get finer control on bicarb conductivity. also the dead band is different. need to figure out whether we can combine or not

This code looks very similar to bicarb code above. Do we really need separate handling?

This code looks very similar to bicarb code above. Do we really need separate handling?

Remove extra blank line.

Remove extra blank line.

Add unit of measure to definition name (e.g. _PSI). Should be done for all definitions w/ a unit of measure to prevent unit errors.

Add unit of measure to definition name (e.g. _PSI). Should be done for all definitions w/ a unit of measure to prevent unit errors.

LEAHI-DD-FIRMWARE-LDT-3980_Dialysate Composition - FW Implementation (Dry Bicarbonate) - 1/1:
LEAHI-DD-FIRMWARE-LDT-3980_Dialysate Composition - FW Implementation (Dry Bicarbonate) - 1/1:
LEAHI-DD-FIRMWARE-LDT-2004_Dialysate Composition - DD
LEAHI-DD-FIRMWARE-LDT-2004_Dialysate Composition - DD
LEAHI-DD-FIRMWARE-LDT-2004_Dialysate Composition - DD
LEAHI-DD-FIRMWARE-LDT-2004_Dialysate Composition - DD
Its static function and not used by external API to pass the wrong id.

Its static function and not used by external API to pass the wrong id.

0.0F is made by Sean.

0.0F is made by Sean.

Same comment applies here (removing resetPIController).

Same comment applies here (removing resetPIController).

I think we can remove the balancing chamber fill in progress check for Opening D80 valve, as this does not impact the balancing chamber fill process.

I think we can remove the balancing chamber fill in progress check for Opening D80 valve, as this does not impact the balancing chamber fill process.

Maintain space after and before parenthesis across the file.

Maintain space after and before parenthesis across the file.

Do we need to add another else statement and address an alarm if wrong mixId is passed? or Keep only else part instead of else if ( ACID_MIX_ID == mixID) check?? this change may be needed across th...

Do we need to add another else statement and address an alarm if wrong mixId is passed? or Keep only else part instead of else if ( ACID_MIX_ID == mixID) check?? this change may be needed across the file.

Remove this comment.

Remove this comment.

You may use getF32OverrideValue for consistency (across the file).

You may use getF32OverrideValue for consistency (across the file).