What's the point of this line if you overwrite the lvl_sen value in the next 2 lines? I think the second lvl_sen is not needed, the one without the offset.

What's the point of this line if you overwrite the lvl_sen value in the next 2 lines?
I think the second lvl_sen is not needed, the one without the offset.

LAST_DD_LEVL is in FW not in dialin. In dialin, you use the Num_of_levels, so this one is correct here. Example: In this case Num_of_levels is 4 (D6,D63,D98,D46), if in the override adding 4 to the...

LAST_DD_LEVL is in FW not in dialin.
In dialin, you use the Num_of_levels, so this one is correct here.
Example:
In this case Num_of_levels is 4 (D6,D63,D98,D46), if in the override adding 4 to the fp level sensor, which is 0, the payload will provide 4, which will be between the FIRST_FP_LEVL and LAST_FP_LEVL FW value as P25 index is 4.

Why increment the number when you didn't add a new sensor?

Why increment the number when you didn't add a new sensor?

This should be LAST_DD_LEVL not NUM_OF_LEVELS if DD and FP levels are in same enum

This should be LAST_DD_LEVL not NUM_OF_LEVELS if DD and FP levels are in same enum

LEVEL_OFFSET?

LEVEL_OFFSET?

LDT-3984 fix for conductivie level sensors d46, d63,d98

Should this be an enumeration somewhere?

Should this be an enumeration somewhere?

LDT-3984 conductive level sensor dialin implementation

LEAHI-DIALIN-LDT-4161_Post-Treatment - SW - 01 - TxReview - R&I
LEAHI-DIALIN-LDT-4161_Post-Treatment - SW - 01 - TxReview - R&I
LEAHI-DIALIN-LDT-3984_(DD, SW) D46 reads 0 with and without liquid in chamber
LEAHI-DIALIN-LDT-3984_(DD, SW) D46 reads 0 with and without liquid in chamber
LDT-3984 aligning D46 with DD firmware

Merge branch 'feature/LDT-3855-sw-enhc-general-alarm' into staging

Merge branch 'feature/LDT-3656-New-Design-CreateRx' into staging

    • -44
    • +59
    /leahi_dialin/ui/td_messaging.py
Merge remote-tracking branch 'origin/staging' into feature/LDT-3855-sw-enhc-general-alarm

Merge branch 'feature/LDT-3757-recirculate' into staging

Merge branch 'feature/LDT-3834-sw-misc-enhc-stop-state' into staging