Done.
remove extra blank line
BiCarb related state machines will be called in pre-gen, gen dialysate and post-gen modes. currently its kept here, so any modes can set a flag to initiate the required operations.
Did not end up needing to do this. Updated model and moved to c++
updated
moved model to the view class
removed
no. updated to remove
fixed
reverted lines
updated both this and advanced views
reverted
updated to have all fonts of the containers to use same variable
added
Use macro for 0.0F
Please remove RTDReadCnt, per HDD.
Bicarb Chamber Low level sensor
Please remove it.
Implement the filter logics using these variables.
Please add the valve validations in the valves.c
Added
Added.
Changed
Removed from .c and .h files
Updated
I need to check with Noe and I will update you by EOD
I will update you by EOD
This should be D98 function.
50
Please add D99 temp average.
fresh dialysate temperature?
freshDialysatePressurePostUltrafilter (instead of TMP)?
D98?
Please add up bicarb chamber level based on the D63 and D98 level sensor combinations?
RTDReadCnt -> Should be defined here?
Does this function need to support the rest of the level sensors? (e.g. P25, D63, D98?)
Don't include an Alarms detail if there are no alarms.
Add blank line before separator.
Got a suggestion or an issue? Discuss it on the Atlassian Community or visit Atlassian Support for other options.