dd-firmware

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
LDT-3352 resolving merge conflicts

    • -4
    • +1
    /firmware/App/Modes/ModeGenDialysate.c
Looks like we are doing a lot of the dialysate temp control in Gen Dialysate mode. Do we not do any heating in PreGen mode? Even if all dialysate heating is done in this mode, should we consider cr...

Looks like we are doing a lot of the dialysate temp control in Gen Dialysate mode. Do we not do any heating in PreGen mode?
Even if all dialysate heating is done in this mode, should we consider creating a service controller (like BC controller) to manage this complex temperature control?

Resolve merge conflicts.

Resolve merge conflicts.

LDT-3372: Reduced warnings in relation to teensy driver. Removed unused functions. Fixed overrides

    • -131
    • +28
    /firmware/App/Drivers/ConductivityTeensy.c
LDT-3344 addressed code review comments

    • -8
    • +119
    /firmware/App/Controllers/RinsePump.c
LDT-2004: updates to v2 driver. changed data types for cal coefficents to f32. Implemented calculated resistance, temp, and conductivity based on raw data from sensor.

    • -21
    • +129
    /firmware/App/Drivers/ConductivitySensors.c
    • -17
    • +8
    /firmware/App/Monitors/Conductivity.c
LDT-3223-dialysate-flow-rate---dd---mode-tel-dev-test : Added testcases for DialysatePump file

Bamboo Commit: Updated DDCommon.h with build versions from Bamboo.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

    • -3
    • +3
    /firmware/App/Drivers/BloodLeakDriver.c
    • -3
    • +3
    /firmware/App/Drivers/BloodLeakDriver.h
Merged LDT-2030 blood leak refactor code

For B1.0/B1.9 - Where Diener silencer 2000 pumps, this macro needs to be enabled.

For B1.0/B1.9 - Where Diener silencer 2000 pumps, this macro needs to be enabled.

There are recent changes done on FPGA register in staging for B1.0/B1.9. please merge the latest changes here.

There are recent changes done on FPGA register in staging for B1.0/B1.9. please merge the latest changes here.

Merge branch 'staging' into LDT-2030-blood-leak-refactor---dd

Bamboo Commit: Updated DDCommon.h with build versions from Bamboo.

Merge branch 'LDT-2030-blood-leak-refactor---dd' into develop

Add valve details per HDD (may be 2.0 valves. as we remove beta 1/1.9 in sometime)

Add valve details per HDD (may be 2.0 valves. as we remove beta 1/1.9 in sometime)

add new valves.

add new valves.

Can we follow the same terminology as DD, starting a variable with 'fpga...'. for all IOFP variables?

Can we follow the same terminology as DD, starting a variable with 'fpga...'. for all IOFP variables?

typo - used.

typo - used.

Add as HDF hallsensor. start the comment with capital letter R.

Add as HDF hallsensor. start the comment with capital letter R.

#define 0U across the file.

#define 0U across the file.

Instead of generic function, recommend adding a specific function something similar to "getACSwitchStatus' or 'getLeaskSensorStatus' and return the values ( ex: leak sensor Enum states?). Do we nee...

Instead of generic function, recommend adding a specific function something similar to "getACSwitchStatus' or 'getLeaskSensorStatus' and return the values ( ex: leak sensor Enum states?). Do we need to add the application code as well now, say if leak is found, trigger an alarm?

either typecast pwmInCount as U08 while passing to FPGA or declare the variable size as U08 instead of U32.

either typecast pwmInCount as U08 while passing to FPGA or declare the variable size as U08 instead of U32.

Bamboo Commit: Updated DDCommon.h with build versions from Bamboo.

resolved merge conflicts

    • -123
    • +148
    /firmware/App/Monitors/Level.c
Add a variable to convert PWM count to target speed for monitor function.

Add a variable to convert PWM count to target speed for monitor function.