DDCommon.h

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Same as above.

Same as above.

Add an event here, when time out happens. Event data should capture fresh side and spent side fill completion status.

Add an event here, when time out happens. Event data should capture fresh side and spent side fill completion status.

You may need to comment/ remove this condition. the pressure stabilized flag should set only when pressure is met (not on chamber fill request)

You may need to comment/ remove this condition. the pressure stabilized flag should set only when pressure is met (not on chamber fill request)

it shouldn't be the default value

it shouldn't be the default value

Bamboo Commit: Updated DDCommon.h with build versions from Bamboo.

There are multiple target bicarb conductivity calls, getBicarbTargetConductivity, getBicardConductivityPre and getBicarConductivityPost. If you are using getBicarbTargetConductivity for deadband co...

There are multiple target bicarb conductivity calls, getBicarbTargetConductivity, getBicardConductivityPre and getBicarConductivityPost. If you are using getBicarbTargetConductivity for deadband control check, use the same for 'runPIcontroller' function as well.

This should return POST conductivity.

This should return POST conductivity.

getBicarbConductivityPre() -> This function gets the target conductivity from the mixing algorithm. For testing, recommend using defined value and check the behavior.

getBicarbConductivityPre() -> This function gets the target conductivity from the mixing algorithm. For testing, recommend using defined value and check the behavior.

Init PI Controls missing here.

Init PI Controls missing here.

13734.87F

13734.87F

Leave a comment here that for acid control period, 3 cycles of balancing chamber switching time (for ex, Qd :600, 3sec is switching period, 3 * 3 = 9 Sec control interval) and for bicarb control, 5...

Leave a comment here that for acid control period, 3 cycles of balancing chamber switching time (for ex, Qd :600, 3sec is switching period, 3 * 3 = 9 Sec control interval) and for bicarb control, 5 cycles of balancing chamber switching time.

both acid and bicarb mix volume

both acid and bicarb mix volume

Add MULTIPLIER at the end.

Add MULTIPLIER at the end.

Initialize at init function.

Initialize at init function.

Remove sample filter ms and task interval. rename as FILTER_SIZE_MULTIPLIER.

Remove sample filter ms and task interval. rename as FILTER_SIZE_MULTIPLIER.

Lets remove this.

Lets remove this.

LEAHI-DD-FIRMWARE-LDT-4410_Dialysate Composition (Mixing) - FW Implementation - 3/1:
LEAHI-DD-FIRMWARE-LDT-4410_Dialysate Composition (Mixing) - FW Implementation - 3/1:
Bamboo Commit: Updated DDCommon.h with build versions from Bamboo.

Bamboo Commit: Updated DDCommon.h with build versions from Bamboo.

Bamboo Commit: Updated DDCommon.h with build versions from Bamboo.

Bamboo Commit: Updated DDCommon.h with build versions from Bamboo.

Bamboo Commit: Updated DDCommon.h with build versions from Bamboo.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 9 more files in changeset.
Done.

Done.

Done.

Done.

Use ternary to explicitly assign TRUE or FALSE.

Use ternary to explicitly assign TRUE or FALSE.