Yes. I put at the top of this function. Because there are only two states and most of times we are running the fans. I also did not want to monitor the fans every 50ms so I put it here to monitor the fans whenever it is time to control the fans.
I think we can create preliminary #defines now. 0.8 L is likely what we want. Control limit can probably go up to 130. Max is still 140, but not relavant to this line of code.
Once I am close to finishing this story, I will go over the doxygen documentation to make sure it is up to date. Please resolve this comment. I will address all the doxygen comment in the next Heat Disinfection code review.
I know I made this change when helping you troubleshoot drain pump control, but looking at this again, I think we need both options for calculating the error signal (depending on which control is in play). I'm concerned that this change may damage some/all of the other controls.