With that approach, we are likely to have pump/heater off for some time which will allow the dialysate to cool and perhaps cause a temperature alarm when we do start treatment. I think we should have the pump/heater on when door is closed in patient connect states.
This code is identical to the DG implementation. Initially they were only included in debug mode. Seemed like it made sense to move to function as they are only used in the function.
Looks ALARM_PRIORITY_LOW ranks are in 700 to 800 series range. Medium priority rank ranging 400 to 600. Do you want to consider this as low priority or medium priority? if medium priority rank, can we get rank in to 400 to 600 series?