Comment Results

Review Name Created Custom Fields Content
LEAHI-TD-FIRMWARE-LDT-4072-2 04 Aug 2026

use FALSE

LEAHI-TD-FIRMWARE-LDT-4072-2 04 Aug 2026

use FALSE

LEAHI-TD-FIRMWARE-LDT-4033-1 05 Aug 2026

Removed

LEAHI-TD-FIRMWARE-LDT-4063-1 05 Aug 2026

Move this above the test functions.

LEAHI-DD-FIRMWARE-LDT-3975-2 05 Aug 2026

Align the indentation.

LEAHI-TD-FIRMWARE-LDT-4040-1 06 Aug 2026

Why are we doing this in blood prime? Wouldn't it be easier to do this when we're done w/ rinseback & recirc?

LEAHI-TD-FIRMWARE-LDT-4063-1 31 Jul 2026

add safetyShutdownSelfTestStatus to outputs

LEAHI-TD-FIRMWARE-LDT-4040-1 06 Aug 2026

I thought Eliza said they wanted to do blood sitting timer even if blood was rinsed back (still a little blood in line).

LEAHI-TD-FIRMWARE-LDT-4063-1 04 Aug 2026

When we send any message to UI, usually add the message id in the doxygen header like this.

  • @details \b Message \b Sent: MSG_ID_TD_RESP_TREATMENT_PARAMS_TO_VALIDATE


Please refer to the function sendTreatmentParamsResponse()

LEAHI-TD-FIRMWARE-LDT-4040-1 04 Aug 2026

those are initialized in resetTreatmentRecircFlags() and that is called in init

LEAHI-TESTSUITES-LDT-3447-1 04 Aug 2026

Revert the staging test script and using the different branch

LEAHI-TD-FIRMWARE-LDT-4031-1 30 Jul 2026

Initialize this in the init function instead of here.

LEAHI-TD-FIRMWARE-LDT-4040-1 23 Jul 2026

Add @message MSG_ID_TD_RECIRCULATE_RESPONSE

LEAHI-TD-FIRMWARE-LDT-4072-2 05 Aug 2026

align comments

LEAHI-TD-FIRMWARE-LDT-4033-1 05 Aug 2026

Fixed

LEAHI-TD-FIRMWARE-LDT-4040-1 05 Aug 2026

Update function name: handleRinsebackAdditionalState

LEAHI-TD-FIRMWARE-LDT-4033-1 05 Aug 2026

Inputs : fpgaSensorReadings
Outputs: none

LEAHI-TD-FIRMWARE-LDT-4033-1 05 Aug 2026

Inputs : fpgaSensorReadings
Outputs: none

LEAHI-TD-FIRMWARE-LDT-4040-1 05 Aug 2026

Add cumulativeBloodPrimeVolume_mL to ouputs.

Optional: While you are updating, can you please also add bloodPrimeRampFlowRate_mL_min and bloodPrimeState to the inputs?

LEAHI-DD-FIRMWARE-LDT-3975-2 05 Aug 2026

Aling the indentation of value field.

LEAHI-DD-FIRMWARE-LDT-3975-2 05 Aug 2026

Space before and after paranthesis.