leahi-common

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
LDT-4612 mixing control updates aligning comments

LDT-4612 mixing control updates

Depressurize?

Depressurize?

please remove extra line

please remove extra line

Output is none

Output is none

Please add message id.

Please add message id.

Fix the typo.

Fix the typo.

Output is none

Output is none

Message id MSG_ID_OFF_BUTTON_PRESS_REQUEST looks wrong

Message id MSG_ID_OFF_BUTTON_PRESS_REQUEST looks wrong

moved to general task

moved to general task

LDT-4612 depressurise state machine implementation

I think we should accept this as long as we're in pre-tx: install state. User might confirm before auto-load finishes and that's ok.

I think we should accept this as long as we're in pre-tx: install state. User might confirm before auto-load finishes and that's ok.

Change "SetupConnection" to "SetupTubingSetConnection"

Change "SetupConnection" to "SetupTubingSetConnection"

Recommend changing "SETUP_CONNECTION" to "SETUP_TUBING_SET_CONNECTIONS" in these 2 msg ID names.

Recommend changing "SETUP_CONNECTION" to "SETUP_TUBING_SET_CONNECTIONS" in these 2 msg ID names.

We need to keep D64 opened after supply to maintain the supply to bicarb green flow filter

We need to keep D64 opened after supply to maintain the supply to bicarb green flow filter

Include "BPModule.h".

Include "BPModule.h".

BPModule is just one of the controllers - move this up into the controllers section.

BPModule is just one of the controllers - move this up into the controllers section.

We should not comment this out.

We should not comment this out.

It's still here. Do we need to initialize it here too or is it sufficient to initialize once from main at startup?

It's still here. Do we need to initialize it here too or is it sufficient to initialize once from main at startup?

We are already executing this from TaskGeneral.c. Remove from here.

We are already executing this from TaskGeneral.c. Remove from here.

Monitor is mis-spelled. Should be Control anyway. Move prior to execAlarmMgmt() above.

Monitor is mis-spelled. Should be Control anyway. Move prior to execAlarmMgmt() above.

I think all flags are outputs (not just adult).

I think all flags are outputs (not just adult).

These look like outputs to me (not inputs).

These look like outputs to me (not inputs).

Updated alarm and msgdefs per review

LEAHI-DD-FIRMWARE-LDT-4409_Dialysate Composition (Dry Bicarbonate) - FW Implementation - 3/1:
LEAHI-DD-FIRMWARE-LDT-4409_Dialysate Composition (Dry Bicarbonate) - FW Implementation - 3/1:
Merge fixes

Merge branch 'staging' into LDT-3975-water-treatment-and-degassing--

Can we restore this code?

Can we restore this code?

Remove FANS test.

Remove FANS test.

Remove.

Remove.