Remove extra blank line.

Remove extra blank line.

Should we create a structure for these moving average filters and make an array of them based on enum of temp sensors? It looks kind of messy having separate handling for each sensor - can we conso...

Should we create a structure for these moving average filters and make an array of them based on enum of temp sensors?
It looks kind of messy having separate handling for each sensor - can we consolidate some of this code?

Should we maybe just have a constant array of sample counts based on enum of temp sensors?

Should we maybe just have a constant array of sample counts based on enum of temp sensors?

Should min and max be properties of the controller instead of global (assumes all controllers will want same min/max)? Controller properties already include a min and max field - can we use those?

Should min and max be properties of the controller instead of global (assumes all controllers will want same min/max)?
Controller properties already include a min and max field - can we use those?

Remove blank line.

Remove blank line.

Done.

Done.

Done.

Done.

mis aligned ?

mis aligned ?

mis aligned ?

mis aligned ?

need a brief description about this feature and how it is implemented/logic/math etc - also comment on the previous logic

need a brief description about this feature and how it is implemented/logic/math etc - also comment on the previous logic

its a reverse logic - is it OK to make positive instead of not defined ?

its a reverse logic - is it OK to make positive instead of not defined ?

These don't need to be initialized.

These don't need to be initialized.

LEAHI-DD-FIRMWARE-LDT-1477_Ultrafiltration - DD - 04: DEV - Feature Implementation
LEAHI-DD-FIRMWARE-LDT-1477_Ultrafiltration - DD - 04: DEV - Feature Implementation
Looks like we are doing a lot of the dialysate temp control in Gen Dialysate mode. Do we not do any heating in PreGen mode? Even if all dialysate heating is done in this mode, should we consider cr...

Looks like we are doing a lot of the dialysate temp control in Gen Dialysate mode. Do we not do any heating in PreGen mode?
Even if all dialysate heating is done in this mode, should we consider creating a service controller (like BC controller) to manage this complex temperature control?

Resolve merge conflicts.

Resolve merge conflicts.

Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 1 more file in changeset.
Add blank line between declarations and rest of code.

Add blank line between declarations and rest of code.

Add comment to right listing which math functions we're using.

Add comment to right listing which math functions we're using.

LEAHI-DD-FIRMWARE-LDT-3480_Improper match on IOFP temperature sensors
LEAHI-DD-FIRMWARE-LDT-3480_Improper match on IOFP temperature sensors
LDT-3480 updated IOFP temp order per category

  1. … 1 more file in changeset.
LEAHI-DD-FIRMWARE-LDT-3352_DD - revisit D3 level control logic
LEAHI-DD-FIRMWARE-LDT-3352_DD - revisit D3 level control logic
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 18 more files in changeset.
resolved merge conflicts

  1. … 6 more files in changeset.
Added

Added

An alarm is triggered in this function. Add to header.

An alarm is triggered in this function. Add to header.

Updated

Updated