TestSupport.h

Clone Tools
  • last updated a few minutes ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
I would keep blank line between library includes and regular includes.

I would keep blank line between library includes and regular includes.

We can't have 2 alarms with ID 196. The enum only has the empty saline bag alarm. Looks like alarm 197 is duplicate of this 196 alarm you've added. I think you should just remove this new alarm 196...

We can't have 2 alarms with ID 196. The enum only has the empty saline bag alarm. Looks like alarm 197 is duplicate of this 196 alarm you've added.
I think you should just remove this new alarm 196 that you've added here.

Fix merge conflict stuff here.

Fix merge conflict stuff here.

Fix merge conflict stuff here.

Fix merge conflict stuff here.

LEAHI-DD-FIRMWARE-LDT-4873_(Request) Add config to disable conductive level sensor fault
LEAHI-DD-FIRMWARE-LDT-4873_(Request) Add config to disable conductive level sensor fault
LDT-4409 enable drybicart by default and support for WCID1 pressure profile

Bubble status, Low and high level level sensors.

Bubble status, Low and high level level sensors.

Input : Blood pump door status?

Input : Blood pump door status?

To be restored?

To be restored?

Are these being initialized in the init function?

Are these being initialized in the init function?

Need to resolve the merge issues?

Need to resolve the merge issues?

LEAHI-DD-FIRMWARE-LDT-4409_Dialysate Composition (Dry Bicarbonate) - FW Implementation - 3/1:
LEAHI-DD-FIRMWARE-LDT-4409_Dialysate Composition (Dry Bicarbonate) - FW Implementation - 3/1:
Bamboo Commit: Updated the Copyright section and replaced tabs with 4 spaces

  1. … 2 more files in changeset.
fixed thanks

fixed thanks

Should we change msg ID from speed to rate?

Should we change msg ID from speed to rate?

fixed thanks

fixed thanks

its a speed in rpm. updated function to convert target rate in ml/min to target speed in rpm and use that for the difference.

its a speed in rpm. updated function to convert target rate in ml/min to target speed in rpm and use that for the difference.

fixed thanks

fixed thanks

this is a set speed in rpm

this is a set speed in rpm

this is speed in rpm

this is speed in rpm

fixed thanks

fixed thanks