Check that Dialin has logged in before acting on a Dialin request (TRUE == isTestingActivated()).
RESOLVED in CODE WALKTHROUGH.
Check editor settings. Tabs should be set to 4 spaces.
RESOLVED in CODE WALKTRHOUGH.
Addressed.
Remove extra line.
This will be addressed once we were working on deciding the number of rinses.
Done.
Bamboo will do the copyrights. It is not turned on right now but it will automatically.
Renamed roPumpSpeed to roPumpFlushBubblesSpeed and updated comments
Updated.
Fixed. Thanks!
Done. Should we remove this build flag all together since no V2 system available anymore?
Missing brief.
Added param: testValue, baseValue, percentFactor
Should this be in an else below?
It doesn't appear to build for me.
Can leave uninitialized here. It is initialized in init function below.
Sent Dara email asking him to put those numbers in the DG calibration record.
RESOLVED.
Could we move these fields outside of the for loop? They dont need to be initialized NUM_OF_FANS_NAMES times.
Bolus rate is fixed now so no range check is needed. Keep check for home though.
RESOVLED IN CODE WALKTHROUGH.
The Arterial and Venous sliders are RangeSlider and are handled differently. The extra code is removed.
Mode 4 for the HD is Treatment parameters. Every time this is called the UI will flicker to the Patient ID screen then back to Home. Is that what we want?