leahi-common

Clone Tools
  • last updated a few seconds ago
Constraints
Constraints: committers
 
Constraints: files
Constraints: dates
Merge branch 'LDT-3978-switch-chamber-f-conductive-lev' of bitbucket.org:diality-cloud/leahi-common into LDT-3978-switch-chamber-f-conductive-lev

Merge branch 'LDT-3978-switch-chamber-f-conductive-lev' of bitbucket.org:diality-cloud/leahi-common into LDT-3978-switch-chamber-f-conductive-lev

LDT-3978 conductive level sensor implementation

LDT-3978 conductive level sensor implementation for chamber F fill

LDT-3735: Added Alarm IDs for NVM module

Merge branch 'develop' of bitbucket.org:diality-cloud/leahi-common into develop

LDT-1886: Added invalid payload rej reason

instead of 0, we can use the level Enum 'LVL_STATE_EMPTY'?

instead of 0, we can use the level Enum 'LVL_STATE_EMPTY'?

lower? and change the D98 level sensor comment as well.

lower? and change the D98 level sensor comment as well.

Remove empty line

Remove empty line

Space before LVL_STATE_LOW

Space before LVL_STATE_LOW

space inside array index.

space inside array index.

Space inside parenthesis.

Space inside parenthesis.

Change this to LEVEL_T enum.

Change this to LEVEL_T enum.

Is this macro still being used? if not, delete it.

Is this macro still being used? if not, delete it.

LDT-2240-disposable-tube-set-self-test

Removed

Removed

Removed

Removed

updated

updated

added

added

Added

Added

Removed

Removed

Updated

Updated

Merge branch 'staging' of bitbucket.org:diality-cloud/leahi-common into LDT-2240-disposable-tube-set-self-test

LEAHI-DD-FIRMWARE-LDT-3978_Switch Chamber F conductive level sensors to broadcast 1 as sensing liquid
LEAHI-DD-FIRMWARE-LDT-3978_Switch Chamber F conductive level sensors to broadcast 1 as sensing liquid
LDT-2240-disposable-tube-set-self-test

Not necessary

Not necessary

This command should be given once on way out of prior state instead of every 50ms while in this state. We also need to zero the timer counter at same time so the timeout check below works.

This command should be given once on way out of prior state instead of every 50ms while in this state.
We also need to zero the timer counter at same time so the timeout check below works.