It appears that we are checking string fields of the ethernet driver and not performing a POST for the driver, is that being performed elsewhere ?, for example ethtool or similar
We need to display the last Saline event before we switch back.
Due to how we exit this state machine, we call transitionToDialysis() if we go to SALINE_BOLUS_STATE_MAX_DELIVERED which will update the substate back to the current UF state.
Do we still need this condition? Door closed requirement may have been released in prior state. Doesn't seem like this is truly needed to move on from here.
I feel like this whole state should be removed. We are no longer doing de-prime. Removing state will mean Dialin must be updated to reflect new state list.
I think we should be checking alarm condition instead of alarm active. Temperature may have come back into range (condition cleared) but user hasn't resumed to clear alarm (maybe stepped away for a minute).
If not done already, we will need to make changes to some of this in HalCoGEN and there may be some of these function calls here that are not needed for RO.
You are running release
CR4.8.14
FE4.8.14
(20240111091859 2024-01-11 09:20),
please report your release number when reporting bugs.