I don't think you are doing a double buffer. Remove this note.
Do we need an end of group marker down here?
TODO
Rename like DRAIN_BICART_PUMP_SPEED. update comment as cartridge drains at maximum speed.
<Prefix @ for advanced mode>
Updated
Updated.
BICARB_CHAMBER_PERIODIC_FILL_TIME
can we give the back buttons a more descirptive name like o_SettingsBase_backButton_UserMode
An alarm is triggered in this function. Add to header.
Remove if not necessary.
I don't see the updates.
Thanks for the update. Please update the styles as well.
Please rename as DRY_BICART_FILL_VENT_TIME_MS
Now I am using application_init.update_password_back_to_default_password() function copy password from system.dflt to system.conf as per comment
Add def aut(name, *args): """ Joins the executable name and argument into a single command string """ return " ".join((name, *args)) function in utility.py and using it main script startApplication(utility.aut("leahi", "-E")) as per comment
Removed duplicate objects
get more specific on opmode and parameter set add TD -> SW -> UI
go UI -> SW - TD and vice versa and update self test to do same
change to 250
add TODO comment i added for my simulation
always show sub pages
fixed for both _TO_PWM macros
fixed.
updated to alwas who sub steps
updated
Fixed it
remove empty line.
Remove this macro.
our standard overrides for conductivity values is covered in the monitor. Set function for conductivity has been implemented.
Get new message IDs from Dara.
Pre-Treatment
Remove wait retry and pre-tx participants
should this be MODE_TREA ?
renamed
Please sync all of these with Firmware order of definitions.
We should use the #defines for .override assignments (i.e. OVERRIDE_RESET or OVERRIDE_KEY).
Remove extra blank line.
RESOLVED in CODE WALKTHROUGH
These values should be retained for Beta 1.0
Change id to _timeoutText.
Got a suggestion or an issue? Discuss it on the Atlassian Community or visit Atlassian Support for other options.