There may be a need to run these pumps in forward or reverse direction (heat disinfect mode). And this would not be tied to RO flow rate - we would want to set an independent flow. May also want to control the pumps independently in this mode.
Please change it to "HDAccelData" as you have stated in the infoText function. this should be exactly the text user may search for in the log file to find the logged message.
We need to set fpga error to 0 to bypass incorrect error count from FPGA. Otherwise, there will be temperature sensor fault alarm sending to HD once DG is running.
Override handlers (and really any handlers for messages that would only come from Dialin) should have the word Test in the name to help distinguish Dialin test functions from other f/w functions.
This folder contains treatment related information such as parameter ranges as well as paired BLE devices. Would you prefer we store the bluetooth devices in a separate folder called bluetooth instead?