Its probably not necessary. i want it as a fallback state for alarms when testing later this week instead of faulting where i can. I'll probably remove it once testing is complete.
Please let the following calculation of the overlap do the separation. Otherwise will completely overlap and eventually displays one value which would be confusing. And also creates a position jump.
Yes, i will remove the Retangle and set the color as we did before, with the change that the Slider's color for active will be a SliderHighlightColor / Orange instead of the light blue
This check to see if it's time to do reprime should be moved to the handler for dialysis: ultrafiltration state (function just below this one) - because we would not want to start a reprime in the middle of a saline bolus or a reprime and because we should only be incrementing the timer counter during actual treatment (when treatment time is not paused).