Thanks for catching that. Any subsystem can send the AlarmCleared, and that is why the Src is DPNDS (depends). I changed the source to Any, and this is an Alarm Broadcast message.
Want Gen Dialysate mode state machine to handle the mode change - so request function should set a flag that the state machine will act on in next pass.
Merge branch 'DEN-16690-UI-MG-S117' into DEN-16839-UI-BN-Whitney-1B1 c88f546ad09c2d6c0222d8e694a9814e37a424ea The follow up CR would be http://devapps.diality.us:8060/cru/CLOUDSYNC-DEN-16839-1 Will be closed.