LEAHI-DD-FIRMWARE-LDT-2004-3 22 Dec 2025

Done.

LEAHI-DD-FIRMWARE-LDT-2004-3 22 Dec 2025

We will move towards 200 RPM being the minimum speed with updated Maxon controller. So, there is no need to maintain two different minimum RPM going forward.

LEAHI-DD-FIRMWARE-LDT-2004-3 22 Dec 2025

Done.

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

remove extra blank line

LEAHI-DD-FIRMWARE-LDT-2004-3 22 Dec 2025

BiCarb related state machines will be called in pre-gen, gen dialysate and post-gen modes. currently its kept here, so any modes can set a flag to initiate the required operations.

LEAHI-APPLICATION-LDT-2825-1 23 Dec 2025

Did not end up needing to do this. Updated model and moved to c++

LEAHI-APPLICATION-LDT-2825-1 23 Dec 2025

updated

LEAHI-APPLICATION-LDT-2825-1 23 Dec 2025

updated

LEAHI-APPLICATION-LDT-2825-1 23 Dec 2025

moved model to the view class

LEAHI-APPLICATION-LDT-2825-1 23 Dec 2025

updated

LEAHI-APPLICATION-LDT-2825-1 23 Dec 2025

removed

LEAHI-APPLICATION-LDT-2825-1 23 Dec 2025

no. updated to remove

LEAHI-APPLICATION-LDT-2825-1 23 Dec 2025

updated

LEAHI-APPLICATION-LDT-2825-1 23 Dec 2025

fixed

LEAHI-APPLICATION-LDT-2825-1 23 Dec 2025

reverted lines

LEAHI-APPLICATION-LDT-2825-1 23 Dec 2025

updated both this and advanced views

LEAHI-APPLICATION-LDT-2825-1 23 Dec 2025

reverted

LEAHI-APPLICATION-LDT-2825-1 23 Dec 2025

updated to have all fonts of the containers to use same variable

LEAHI-APPLICATION-LDT-2825-1 23 Dec 2025

added

LEAHI-APPLICATION-LDT-2825-1 23 Dec 2025

updated

LEAHI-APPLICATION-LDT-2825-1 23 Dec 2025

added

LEAHI-DD-FIRMWARE-LDT-2185-1 22 Dec 2025

Use macro for 0.0F

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

Please remove RTDReadCnt, per HDD.

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

Bicarb Chamber Low level sensor

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

Please remove it.

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

Implement the filter logics using these variables.

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

Please add the valve validations in the valves.c

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

Added

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

Added.

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

Added.

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

Added

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

Added

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

Changed

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

Changed

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

Removed from .c and .h files

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

Updated

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

Updated

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

Updated

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

I need to check with Noe and I will update you by EOD

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

I will update you by EOD

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

This should be D98 function.

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

50

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

Please add D99 temp average.

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

fresh dialysate temperature?

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

freshDialysatePressurePostUltrafilter (instead of TMP)?

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

D98?

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

Please add up bicarb chamber level based on the D63 and D98 level sensor combinations?

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

RTDReadCnt -> Should be defined here?

LEAHI-DD-FIRMWARE-LDT-2998-1 30 Dec 2025

Does this function need to support the rest of the level sensors? (e.g. P25, D63, D98?)

LEAHI-DD-FIRMWARE-LDT-2185-1 18 Dec 2025

Don't include an Alarms detail if there are no alarms.